The perfect GC for many applications that saves space without compromise on functionality and performance. Capacity for 2 injectors , 1 GC detector plus a mass spectrometer makes the SCION 436 a perfect match for your analyses. A 9 inch touchpad provides direct access to all GC functions and basic tuning.
- Compact footprint
- Fast ramping oven (170°C/minute)
- Flexible 2 channel architecture
- 2 Injectors, 1 Detectors + MS
- Electronic flow control (EFC), 3 channels 0.001psi resolution
- Large full colour screen, multiple languages supported
- Full automation capabilities through CompassCDS software and auto sampler
Dimensions and Weights
Height*: 57 cm (22.5 in.),
Width: 32 cm (13.0 in.),
Depth: 61 cm (24.0 in.)
Weight*: 26.8 kg (59lb)
* Typical values
Environmental Conditions
Operating temperatures: 10°C to 40°C.
Operating humidity (relative): 5 % to 95 %
Line voltage requirements: 120 V, 230 V (±10 % nominal)
Column Oven
Dimensions: 23 cm (w) x 11 cm (d) x 28 cm (h)
Temperature range:
- Ambient: +4°C to 450°C
- Liquid N2: -100°C to 450°C
- Liquid CO2: -60°C to 450°C
Temperature program ramps/holds: 24/25
Maximum temperature ramp rate: 170°C/min for all voltages
Cool down rate: 400 °C to 50 °C in 4.5 minutes
Temperature set-point resolution: 0.1 °C
Ambient temperature reject <0.01°C change in oven for 1°C change in ambient temp.
Retention Time Repeatability <0.008% or < 0.0008 min, based on n-Pentadecane under temperature programming conditions.
Area repeatability < 1% RSD
General Specifications
Up to 3 EFC modules total, injector, detector and auxiliary
Optional backflush
GC Control:
External events (digital output):
- 8 standard
- 8 optional, total 16
Max number of timed events: 30 #
Heated zones:
- Standard 5
Two power outlets 24V (1A max. each)
Methods:
- Maximum stored internal methods: 50 (max. 30 alphanumeric characters)
Logging:
- Run log file (stored with the chromatogram when using CompassCDS)
- Error log file
Local Display:
- TFT full color screen
- WVGA resolution (800 x 480); size 23 cm (9”)
Local Control:
- Touch screen
- Hard Keys
Languages:
- English, German, French, Spanish, Italian, Portuguese, Cyrillic, Kanji, Chinese (standard and traditional), Thai, Korean
- and Dutch
Local automation:
- Method lines: 25
- Modes: (Infinite looping, Dual and duplicate injection)
Communication
Ethernet: Protocol: TCP/IP Data rate: 100 Mbps
Control:
- GC control and method parameters
- Analog output (optional): (Number of channels: 1, Time programmable steps: 30)
- Output software selectable (set individual): (0-1 V (default), 0-10 V)
Synchronization signals with other devices and data systems:
Ready in and out
Start in and out
Data Handling and System Control:
CompassCDS Chromatography Data System
Certifications
- CSA: C22.2 61010-1,UL 61010-1, IEC: 61010-1
- EMC: 47 CFR part 15, ANSI C63.4, EN 61326
- ATEX Directive 94/9/EC
Injector Options
Maximum injectors: two, operating concurrently
Pneumatics: Electronic Flow Control (EFC)
Injector types:
- S/SL Split/Splitless injector*
- PTV Programmable Temperature Vaporizing*
- COC Cold On-Column injector*
- Flash injector
- PWC Packed/ Wide bore On-Column injector
* including septum purge
S/SL Split/Splitless Injector
Pressure range: 0-150 psi
Total flow:
- 500 mL/min for N2/Ar
- 1500 mL/min for He/H2
Maximum temperature: 450°C
Split range: 1-10,000 (column dependent)
Suited for columns:
- Wide bore: (0.53 mm)
- Narrow bore: (0.05 to 0.32 mm
COC Cold On-Column Injector
Pressure range: 0-150 psi
Total Flow: 50 mL/min (Type 23 EFC)
Temperature range:
- Ambient +10 °C to 450 °C using air cooling
- -60 °C to 450 °C using liquid CO2 cooling
- -160 °C to 450 °C using liquid N2 cooling
Maximum temperature: 450°C
Maximum temperature ramp rate: 200°C/min
Temperature ramps/holds: 24/25
Suited for columns:
- Wide bore (0.53 mm)
- Narrow bore
PTV Programmable Temperature
Vaporizing Injector
Pressure range: 0-150 psi
Total flow:
- 500 mL/min for N2/Ar
- 1500 mL/min for He/H2
Temperature range:
- Ambient + 10 °C to 450°C using air cooling
- -160 °C to 450 °C using liquid N2 cooling
- -60 °C to 450 °C using liquid CO2 cooling
Maximum temperature ramp rate: 200°C/min
Temperature ramps/holds: 24/25
Split range: 1-10,000 (column dependent)
Operational capabilities:
- Large volume injection
- Temperature ramped splitless
- Cold on-column
- Split and splitless
Suited for columns:
- Wide bore (0.53 mm)
- Narrow bore (0.05 to 0.32 mm)
- Maximum injection volume: 250 µL (LVI mode)
Flash Injector
Pressure range: 0-150 psi
- Total flow: 50 mL/min (Type 23 EFC)
Maximum temperature: 450°C
Suited for columns:
- Wide bore (0.53 mm)
- Packed (1/8” to 1/4”)
PWOC Packed/Wide-bore On-Column Injector
Pressure range: 0-150 psi
- Total flow: 50 mL/min (Type 23 EFC)
Maximum temperature: 450°C
Suited for columns:
- Wide bore (0.53 mm)
- Packed (1/8” to 1/4”)
Electronic Flow Control: Injectors (EFC)
Module types: 2 injector-specific modules
Pressure: 0.1 % Full Scale
Resolution pressure set points is 0.001psi
Flow sensor accuracy 2% of measured or 0.2% of full scale Flow sensor repeatability 0.5%
Detector Options
Maximum detectors: two, operating concurrently (one of
which is a Single or Triple Quad MS)
Pneumatics: Electronic Flow Control (DEFC)
Detector types:
- FID Flame Ionization Detector
- TCD Thermal Conductivity Detector
- ECD Electron Capture Detector NPD
- NPD Nitrogen-Phosphorus Detector
- PFPD Pulsed Flame Photometric Detector
- PDHID Pulsed Discharge Helium Ionization Detector
- MS Mass Spectrometry (see GC/MS brochure and datasheet)
Note: Data Acquisition Rate: 600Hz for all detectors,
(exception PFPD)
FID Flame Ionization Detector
Maximum temperature: 450 °C
- Detectivity: 1.4 pg C/sec
- Linear dynamic range: 10 7
Flame tip type: ceramic (patented)
Operational quality:
- Flame-out detection
- Auto re-ignition
TCD Thermal Conductivity Detector
Maximum temperature: 450 °C
Detectivity: 300 pg/mL (Butane)
Linear dynamic range: 10 6
Operational quality:
- Filament protection
- Automatic bridge balancing
ECD Electron Capture Detector
Maximum temperature: 450 °C
Detectivity: 7 fg/s Lindane
Linear dynamic range: 10 4
Radioactive source: 63Ni – 15 mCi (555 Mbq)
NPD Nitrogen-Phosphorus Detector
Maximum temperature: 450°C
Detectivity:
N: 100 fg N/sec (Azobenzene) P: 100 fg P/sec (Malathion)
Linear dynamic range:
- N: 10 5
- P: 10 4
PFPD Pulsed Flame Photometric Detect
Photomultiplier tube: S/P
- S/P/N
Maximum temperature: 450 °C
Detectivity:
- S: 1 pg S/sec (S/P tube)
- P: 100 fg P/sec (S/P tube)
- N: 20 pg N/sec (S/P/N tube)
Linear dynamic range:
- S: 10 3
- P: 10 4
- N: 102
- Up to 23 elements can be detected
PDHID Pulsed Discharge Helium Ionization Detector Detectivity:
50 ppb (Methane)
Linear dynamic range: 104
(Methane)
Operational quality:
- Gold plated connections
- Welded column connections
Detectors (DEFC)
Module types: 6 detector-specific modules
Accuracy: ± 7 % set point flow
Resolution: 0.1 or 1 mL/min
Automation Options
8410 Autoinjector
Sample capacity:
- 10 x 2 mL vials
- 6 x 5 mL vials
- 5 x 10 mL vials
Large solvent wash vial: 2 x 120 mL*
Dual and duplicate mode
Internal standard addition
Modes of operation:
- Liquid
- Ambient headspace*
- SPME (Solid Phase Micro Extraction)*
- Sample heating and cooling*
Pre-programmed modes of injection Syringes:
- 1 μL, 2 μL, 5 μL, 10 μL, 100 μL, 250 μL for liquid injection
- SPME
8400 Autosampler
Sample capacity: 100 x 2 mL vials
Large solvent wash vial: 2 x 120
mL* Dual and duplicate mode
Internal standard addition
Modes of operation:
- Liquid
- Ambient headspace*
- SPME*
- Sample heating and cooling*
- Pre-programmed modes of injection
Syringes:
- 1μL, 2 μL,5 μL,10 μL,100 μL,250 μL for liquid injection
- SPME
* Optional
Optionally a CTC-PAL autosampler, including specific options
may be installed.
- Compact footprint
- Fast ramping oven (170°C/minute)
- Flexible 2 channel architecture
- 2 Injectors, 1 Detectors + MS
- Ultimate flexibility/configurability.
- Large capacity, fast ramping oven (150°C/minute)
- Capacity: 3 Injectors, 3 Detectors + MS.
- Full Electronic Flow Control (EFC), up to 21 channels at 0.001psi resolution.
- Innovative lens-free ion path delivers simplifies tuning
- Off axis source and detector increase sensitivity
- Active focusing Q0
- Dual filament source increases uptime
- Compound-based scanning software
- Compact footprint
- Fast ramping oven (170°C/minute)
- Flexible 2 channel architecture
- 2 Injectors, 1 Detectors + MS
- Ultimate flexibility/configurability.
- Large capacity, fast ramping oven (150°C/minute)
- Capacity: 3 Injectors, 3 Detectors + MS.
- Full Electronic Flow Control (EFC), up to 21 channels at 0.001psi resolution.
- Innovative lens-free ion path delivers simplifies tuning
- Off axis source and detector increase sensitivity
- Active focusing Q0
- Dual filament source increases uptime
- Compound-based scanning software